• Ergonomic design provides excellent operating experience
• Large display window allows for easier volume identification
• Pipettes cover volume range from 0.1μL to 10mL
• Easy calibration and maintenance
• Each MicroPette supplied with an individual calibration certificate according to ISO8655
Single-channel Adjustable Volume Pipettes
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error |
Maximum permissible random error |
||
(Inaccuracy) |
(Imprecision) |
|||||
% |
μL |
% |
μL |
|||
0.1-2.5μL |
0.05μL |
2.5μL |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25μL |
3.00% |
0.0375 |
3.00% |
0.0375 |
||
0.25μL |
12.00% |
0.03 |
6.00% |
0.015 |
||
0.5-10μL |
0.1μL |
10μL |
1.00% |
0.1 |
0.80% |
0.08 |
5μL |
1.50% |
0.075 |
1.50% |
0.075 |
||
1μL |
2.50% |
0.025 |
1.50% |
0.015 |
||
2-20μL |
0.5μL |
20μL |
0.90% |
0.18 |
0.40% |
0.08 |
10μL |
1.20% |
0.12 |
1.00% |
0.1 |
||
2μL |
3.00% |
0.06 |
2.00% |
0.04 |
||
5-50μL |
0.5μL |
50μL |
0.60% |
0.3 |
0.30% |
0.15 |
25μL |
0.90% |
0.225 |
0.60% |
0.15 |
||
5μL |
2.00% |
0.1 |
2.00% |
0.1 |
||
10-100μL |
1μL |
100μL |
0.80% |
0.8 |
0.15% |
0.15 |
50μL |
1.00% |
0.5 |
0.40% |
0.2 |
||
10μL |
3.00% |
0.3 |
1.50% |
0.15 |
||
20-200μL |
1μL |
200μL |
0.60% |
1.2 |
0.15% |
0.3 |
100μL |
0.80% |
0.8 |
0.30% |
0.3 |
||
20μL |
3.00% |
0.6 |
1.00% |
0.2 |
||
50-200μL |
1μL |
200μL |
0.60% |
1.2 |
0.15% |
0.3 |
100μL |
0.80% |
0.8 |
0.30% |
0.3 |
||
50μL |
1.00% |
0.5 |
0.40% |
0.2 |
||
100-1000μL |
5μL |
1000μL |
0.60% |
6 |
0.20% |
2 |
500μL |
0.70% |
3.5 |
0.25% |
1.25 |
||
100μL |
2.00% |
2 |
0.70% |
0.7 |
||
200-1000μL |
5μL |
1000μL |
0.60% |
6 |
0.20% |
2 |
500μL |
0.70% |
3.5 |
0.25% |
1.25 |
||
200μL |
0.90% |
1.8 |
0.30% |
0.6 |
||
1000-5000μL |
50μL |
5000μL |
0.50% |
25 |
0.15% |
7.5 |
2500μL |
0.60% |
15 |
0.30% |
7.5 |
||
1000μL |
0.70% |
7 |
0.30% |
3 |
||
2-10mL |
0.1mL |
10mL |
0.60% |
60 |
0.20% |
20 |
5mL |
1.20% |
60 |
0.30% |
15 |
||
2mL |
3.00% |
60 |
0.60% |
12 |