Nucleic Acid Purification System — Auto-Pure 4800
Auto-Pure 4800 automatic nucleic acid extraction and purification instrument is a high-throughput automatic nucleic acid extraction equipment. The instrument is based on the magnetic bead method to extract nucleic acid. During the purification process, the steps of pipetting, lysis, washing, elution, and PCR system set-up are automatically completed in the workstation. One-stop operation, without manual intervention. Can extract 48×1 ml sample system in one run.
Information management
Support external USB scanner, barcode scanning function, automatic scanning function
Pipetting principle
Air displacement pipetting technology
Pipetting range (μL)
1-1000 μL
Liquid detection
Pressure detection: liquid level detection, tip clogging detection and residual liquid detection
Samples per run
1-48
Extraction processing volume
50-1000 μL
Extraction uniformity
CV≤5 %
Single sample consumable
Support
Extraction temperature control
Elution, lysis temperature control: RT~ 120 ℃
Reagent cooling module
Optional, 0 ℃~105 ℃
Temperature control module (Auto-Pure 4800M)
4 ℃~105 ℃
UV sterilization
Separate UV lamp: extraction area, pipetting area and waste container area
Instrument port
USB port
Power supply
220 V, 8 A, 50/60 Hz
Size (WxDxH) mm
1090x700x724
Net weight
130 kg
On Wing Tat Co. Ltd