DESCRIPTION
Measurement Parameters
• BEP-M320: pH, mV, relative mV, ion concentration, temperature
Features
pH
• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers
• Automatic electrode diagnosis shows the pH slope and offset
ORP
• 1 point offset calibration allows to adjust the displayed value to a known standard
• Relative and absolute millivolt modes ensure the reliable ORP measurements
Ion Concentration
• 2 to 5 points calibration, including the 8 concentration points that can be selected
• Direct ion concentration readout simplifies the measurement process
• Automatic electrode diagnosis shows the calibration points and slopes
Other Features
• Automatic temperature compensation ensures accurate readings over the entire range
• Auto-Read function senses and locks the measurement endpoint
• Calibration due alarm prompts user to calibrate the meter regularly
• Auto-Power Off effectively conserves battery life
• Setup menu allows to set the number of calibration points, resolution, stability criteria, temperature unit, date and time, etc.
• Reset function automatically resumes all settings back to the factory defaults
• Expanded memory stores or recalls up to 500 data sets
• USB communication interface is easy to transfer data to PC
• Multi-mode power scheme (batteries, power adapter and USB port) ensures that using
Optional Ion Selective Electrodes
Ammonium (NH4+), bromide (br –), cadmium (Cd2+), calcium (Ca2+), chloride (Cl-), cupric (Cu2+), cyanide (Cn –), fluoride (F –), lodide (I –), lead (Pb2+), nitrate (NO3- ), potassium (K+), silver (Ag+), sodium (Na+), sulphide (S2-) and ammonia (NH3)
SPECIFICATIONS
Model |
BEP-M320 |
|
PH |
Range |
-2.000-20.0OOpH |
Resolution |
0.001,0.01,0.1 pH, selectable |
|
Accuracy |
0.002pH |
|
Calibration Points |
1 to 5 points |
|
pH Buffer Options |
USA, NIST, DIN,2 custom buffers |
|
Ion |
Range |
0.001-19999(depending on the range of ISE) |
Resolution |
0.001,0.01,0.1,1 |
|
Accuracy |
±0.5% FS. (monovalent), ±1% FS. (divalent) |
|
Measurement Units |
ppm, mg/L, mol/L, mmol/L |
|
Calibration Points |
2to 5 points |
|
Calibration Solutions |
0.001,0.01,0.1,1,10,100,1000,10000 |
|
ORP |
mV Range |
±1999.9mV |
Relative mV Range |
±1999.9mV |
|
Resolution |
0.1, 1mV, selectable |
|
Accuracy |
±0.2mV |
|
Calibration Points |
1 point |
|
Temperature |
Range |
0~105°C/32~221°F |
Resolution |
0.1°C/0.1°F |
|
Accuracy |
±0.5°C/±0.9°F |
|
Offset Calibration |
1 point, reading ±10°C |
|
Other Specifications |
temperature Compensation |
0~100°C, manual or automatic |
Stability Criteria |
Low or high |
|
Calibration Due Alarm |
1 to 31 days or off |
|
Slope/Offset Display |
Yes |
|
Hold Function |
Manual or auto-endpoint |
|
Auto-Off |
10, 20 or 30 minutes after the last key is pressed |
|
Memory |
500 data sets |
|
Communication Interface |
USB |
|
Connector |
BNC,3.5mm jack socket |
|
Display |
Custom LCD (80×60mm) |
|
Power Requirements |
3× 1.5V AA batteries or DC5V power adapter |
|
Battery Life |
Approximately 150 hours (Turn off the backlight) |
|
Dimensions |
170(L)×85(W)×30(H)mm |
|
Weight |
300g |
On Wing Tat Co. Ltd